Noise floor and RBW calculator

Type the RBW and noise figure to see how low a spectrum analyzer can see, how long a sweep takes and what a preamp changes.

Noise floor (DANL)—
Noise density—
Sweep time—
Signal to noise—

    How to use the noise floor calculator

    1. Type the RBW your analyzer is set to, or press one of the quick buttons.
    2. Type the analyzer's noise figure. Data sheets often give the DANL instead: noise figure = DANL (in a 1 Hz RBW) + 174. A DANL of −150 dBm/Hz means a noise figure of 24 dB.
    3. Add the span for a sweep time estimate, and the level of the signal you want to see.
    4. Turn on the preamp or add attenuation to see how much they move the floor. Copy the results or share the link.

    Worked example: RBW 10 kHz, noise figure 25 dB

    RBW 100 kHz: −99 dBm RBW 10 kHz: −109 dBm RBW 1 kHz: −119 dBm signal −100 dBm each 10× narrower RBW lowers the noise floor by 10 dB
    The same −100 dBm signal is 9 dB above the floor with a 10 kHz RBW, and hidden with 100 kHz.

    Start with the thermal noise in 1 Hz: −174 dBm. An RBW of 10 kHz is 10,000 Hz wide, and 10 × log10(10,000) = 40 dB, so the filter lets in −174 + 40 = −134 dBm of thermal noise. The analyzer adds its own noise, 25 dB of noise figure, which gives −134 + 25 = −109 dBm. That is the displayed average noise level, or DANL: the flat grassy line at the bottom of the screen.

    A signal at −100 dBm is 9 dB above it, so it shows as a small bump. Switch to a 100 kHz RBW and the floor rises by 10 dB to −99 dBm, and the same signal disappears into the noise. Go down to 1 kHz and the floor falls to −119 dBm, giving the signal 19 dB of room. With a 100 MHz span, the sweep takes about 2.5 × 100,000,000 ÷ 10,000² = 2.5 s at 10 kHz, 25 ms at 100 kHz and 250 s at 1 kHz: four minutes for one sweep.

    The readouts show the same steps. "Noise density" is the floor in 1 Hz (−174 dBm/Hz plus the noise figure), handy for comparing analyzers. "Signal to noise" is how far your signal stands above the floor; below about 3 dB it cannot be told apart from the noise.

    Use cases

    See a weak signal

    You expect a 433.92 MHz key fob at about −110 dBm at your bench, and your analyzer's noise figure is 28 dB. At 100 kHz RBW the floor is −96 dBm, 14 dB too high. The calculator shows that 1 kHz RBW brings it to −116 dBm, 6 dB under the signal. Now you know which setting to use before you start, and you can check the result later in the RF trace viewer.

    Choose RBW vs sweep time

    Scanning the whole 2.4 GHz WiFi band (100 MHz span) at 1 kHz RBW takes about 4 minutes per sweep, which is far too slow to catch a device that sends short bursts. At 30 kHz the sweep takes 0.28 s and the floor is only 15 dB higher. The rule of thumb: pick the widest RBW that still keeps your signal 10 dB above the floor.

    Add a preamp

    A preamp with 20 dB of gain and a 1 dB noise figure in front of an analyzer with a 25 dB noise figure brings the whole chain down to about 6.4 dB, so the floor drops by about 18.5 dB. A 10 dB attenuator does the opposite: it raises the floor by 10 dB, which is the price of protecting the input from a strong signal. The calculator adds them with the Friis formula, so you also see that a preamp with only 10 dB of gain helps much less. For longer chains, use the cascade noise figure calculator.

    Formula

    Displayed average noise level: DANL (dBm) = −174 dBm/Hz + 10 × log10(RBW in Hz) + NF (dB). The −174 comes from kTB: Boltzmann's constant (1.38 × 10−23 J/K) times 290 K times 1 Hz, which is 4 × 10−21 W, or −173.98 dBm.

    Chain noise figure (Friis): F = F1 + (F2 − 1) ÷ G1 + (F3 − 1) ÷ (G1 × G2), with every value as a plain ratio (F = 10NF/10). An attenuator of L dB has a gain of −L dB and a noise figure of L dB.

    Sweep time: t ≈ k × span ÷ (RBW × VBW) when the VBW is narrower than the RBW, otherwise k × span ÷ RBW². k is about 2 to 3 for analog filters. Analyzers with FFT or digital filters sweep much faster at narrow RBW, so treat this as the slow, worst case.

    Limits

    The answer is the average noise power. Most analyzers show noise with log averaging, which reads about 2.5 dB lower, and the noise "grass" itself swings a few dB up and down, so a real screen will look slightly different. The noise figure in a data sheet is often given at one frequency and gets worse near the top of the range, and spurs inside the analyzer can show above the floor. The calculator also assumes a matched 50 Ω input at room temperature. For the other settings that shape the screen, see spectrum analyzer settings, and to change the units, the dBm converter. All RF tools are on the RF tools page.

    FAQ

    What is RBW?

    Resolution bandwidth: the width of the analyzer's filter. Narrower RBW lowers the noise floor and separates close signals.

    How much does RBW change the noise floor?

    10 dB for every 10× change: 100 kHz to 10 kHz lowers it by 10 dB.

    Why does narrow RBW make sweeps slow?

    Sweep time grows with span ÷ RBW², so a 10× narrower RBW is about 100× slower.

    What is −174 dBm/Hz?

    The thermal noise power in 1 Hz at room temperature (kTB).

    Written and tested by the SpectrumAnalyzer.app team · Last checked

    How we tested: Checked against the worked example of the plan: RBW 10 kHz with a 25 dB noise figure gives −109.0 dBm, and each 10× change of RBW moves the floor by 10.0 dB.